MidnightBSD

Advisories for abbott

CVE-2017-12712 HIGH

The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

CVSS 2.0

Severity: HIGH

Problem Type: CWE-287,CWE-287,

Products Affected

Vendor Product Version
abbott anthem_firmware *
abbott accent_mri_firmware *
abbott allure_firmware *
abbott assurity_firmware *
abbott accent_st_firmware *
abbott accent_firmware *
abbott assurity_mri_firmware *
CVE-2017-12714 MEDIUM

Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

CVSS 2.0

Severity: MEDIUM

Problem Type: CWE-920,CWE-920,

Products Affected

Vendor Product Version
abbott anthem_firmware *
abbott accent_mri_firmware *
abbott allure_firmware *
abbott assurity_firmware *
abbott accent_st_firmware *
abbott accent_firmware *
abbott assurity_mri_firmware *
CVE-2017-12716 LOW

Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

CVSS 2.0

Severity: LOW

Problem Type: CWE-311,CWE-319,

Products Affected

Vendor Product Version
abbott anthem_firmware *
abbott accent_mri_firmware *
abbott accent_st_firmware *
abbott accent_firmware *
CVE-2017-5149 MEDIUM

An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical's web site, Merlin.net, are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints.

CVSS 3.x

Source Score Severity Vector Exploitability Impact
nvd@nist.gov 8.9 HIGH CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:L/I:H/A:H 2.2 6.0

CVSS 2.0

Severity: MEDIUM

Problem Type: CWE-476,

Products Affected

Vendor Product Version
abbott merlin@home_firmware *
CVE-2020-8997 MEDIUM

Older generation Abbott FreeStyle Libre sensors allow remote attackers within close proximity to enable write access to memory via a specific NFC unlock command. NOTE: The vulnerability is not present in the FreeStyle Libre 14-day in the U.S (announced in August 2018) and FreeStyle Libre 2 outside the U.S (announced in October 2018).

CVSS 3.x

Source Score Severity Vector Exploitability Impact
nvd@nist.gov 8.8 HIGH CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H 2.8 5.9

CVSS 2.0

Severity: MEDIUM

Problem Type: CWE-787,

Products Affected

Vendor Product Version
abbott freestyle_libre_firmware -
CVE-2023-47262

The startup process and device configurations of the Abbott ID NOW device, before v7.1, can be interrupted and/or modified via physical access to an internal serial port. Direct physical access is required to exploit.

CVSS 3.x

Source Score Severity Vector Exploitability Impact
nvd@nist.gov 5.2 MEDIUM CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:L 0.9 4.2

Products Affected

Vendor Product Version
abbott id_now_firmware *