MidnightBSD

Advisories for hospira

CVE-2014-5400 MEDIUM

The installation component in Hospira MedNet before 6.1 places cleartext credentials in configuration files, which allows local users to obtain sensitive information by reading a file.

CVSS 2.0

Severity: MEDIUM

Problem Type: CWE-260,CWE-200,

Products Affected

Vendor Product Version
hospira mednet *
CVE-2014-5401 HIGH

Hospira MedNet software version 5.8 and prior uses vulnerable versions of the JBoss Enterprise Application Platform software that may allow unauthenticated users to execute arbitrary code on the target system. Hospira has developed a new version of the MedNet software, MedNet 6.1. Existing versions of MedNet can be upgraded to MedNet 6.1.

CVSS 2.0

Severity: HIGH

Problem Type: CWE-94,CWE-94,

Products Affected

Vendor Product Version
hospira mednet *
CVE-2014-5403 MEDIUM

Hospira MedNet before 6.1 uses hardcoded cryptographic keys for protection of data transmission from infusion pumps, which allows remote attackers to obtain sensitive information by sniffing the network.

CVSS 2.0

Severity: MEDIUM

Problem Type: CWE-321,CWE-310,

Products Affected

Vendor Product Version
hospira mednet *
CVE-2014-5405 HIGH

Hospira MedNet before 6.1 uses a hardcoded cleartext password to control SQL database authorization, which allows remote authenticated users to bypass intended access restrictions by leveraging knowledge of this password.

CVSS 2.0

Severity: HIGH

Problem Type: CWE-259,CWE-200,

Products Affected

Vendor Product Version
hospira mednet *
CVE-2014-5406 HIGH

The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.

CVSS 2.0

Severity: HIGH

Problem Type: CWE-345,CWE-345,

Products Affected

Vendor Product Version
hospira lifecare_pcainfusion_firmware *
CVE-2015-1011 MEDIUM

Hospira LifeCare PCA Infusion System before 7.0 has hardcoded credentials, which makes it easier for remote attackers to obtain access via unspecified vectors.

CVSS 2.0

Severity: MEDIUM

Problem Type: CWE-200,

Products Affected

Vendor Product Version
hospira lifecare_pcainfusion_firmware *
CVE-2015-3459 HIGH

The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified commands.

CVSS 2.0

Severity: HIGH

Problem Type: CWE-264,

Products Affected

Vendor Product Version
hospira lifecare_pca3 -
hospira lifecare_pcainfusion_firmware *
hospira lifecare_pca5 -
CVE-2015-3955 HIGH

Stack-based buffer overflow in Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to execute arbitrary code via unspecified vectors.

CVSS 2.0

Severity: HIGH

Problem Type: CWE-119,

Products Affected

Vendor Product Version
hospira lifecare_pcainfusion_firmware *
CVE-2015-3957 MEDIUM

Hospira LifeCare PCA Infusion System before 7.0 stores private keys and certificates, which has unspecified impact and attack vectors.

CVSS 2.0

Severity: MEDIUM

Problem Type: CWE-255,

Products Affected

Vendor Product Version
hospira lifecare_pca3 -
hospira lifecare_pcainfusion_firmware *
hospira lifecare_pca5 -
CVE-2015-3958 HIGH

Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to cause a denial of service (forced manual reboot) via a flood of TCP packets.

CVSS 2.0

Severity: HIGH

Problem Type: CWE-19,

Products Affected

Vendor Product Version
hospira lifecare_pcainfusion_firmware *
CVE-2015-7909 HIGH

Stack-based buffer overflow in Hospira Communication Engine (CE) before 1.2 in LifeCare PCA Infusion System 5.07, Plum A+ Infusion System 13.40, and Plum A+3 Infusion System 13.40 allows remote attackers to cause a denial of service or possibly have unspecified other impact via traffic on TCP port 5000.

CVSS 2.0

Severity: HIGH

Problem Type: CWE-119,

Products Affected

Vendor Product Version
hospira lifecare_pca_infusion_system 5.0.7
hospira communication_engine *