An issue was discovered on Nescomed Multipara Monitor M1000 devices. The device enables an unencrypted TELNET service by default, with a blank password for the admin account. This allows an attacker to gain root access to the device over the local network.
CVSS 3.x
| Source | Score | Severity | Vector | Exploitability | Impact |
|---|---|---|---|---|---|
| nvd@nist.gov | 7.8 | HIGH | CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H | 1.8 | 5.9 |
CVSS 2.0
Severity: HIGH
Problem Type: CWE-287,CWE-319,
Products Affected
| Vendor | Product | Version |
|---|---|---|
| niscomed | m1000_multipara_patient_monitor_firmware | - |
An issue was discovered on Nescomed Multipara Monitor M1000 devices. The physical UART debug port provides a shell, without requiring a password, with complete access.
CVSS 3.x
| Source | Score | Severity | Vector | Exploitability | Impact |
|---|---|---|---|---|---|
| nvd@nist.gov | 6.8 | MEDIUM | CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H | 0.9 | 5.9 |
CVSS 2.0
Severity: HIGH
Problem Type: CWE-306,
Products Affected
| Vendor | Product | Version |
|---|---|---|
| niscomed | m1000_multipara_patient_monitor_firmware | - |
An issue was discovered on Nescomed Multipara Monitor M1000 devices. The internal storage of the underlying Linux system stores data in cleartext, without integrity protection against tampering.
CVSS 3.x
| Source | Score | Severity | Vector | Exploitability | Impact |
|---|---|---|---|---|---|
| nvd@nist.gov | 7.5 | HIGH | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N | 3.9 | 3.6 |
CVSS 2.0
Severity: MEDIUM
Problem Type: CWE-312,
Products Affected
| Vendor | Product | Version |
|---|---|---|
| niscomed | m1000_multipara_patient_monitor_firmware | - |
An issue was discovered on Nescomed Multipara Monitor M1000 devices. The onboard Flash memory stores data in cleartext, without integrity protection against tampering.
CVSS 3.x
| Source | Score | Severity | Vector | Exploitability | Impact |
|---|---|---|---|---|---|
| nvd@nist.gov | 5.5 | MEDIUM | CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:N/A:N | 1.8 | 3.6 |
CVSS 2.0
Severity: LOW
Problem Type: CWE-312,
Products Affected
| Vendor | Product | Version |
|---|---|---|