MidnightBSD

Advisories for pifzer

CVE-2015-3952 MEDIUM

Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

CVSS 2.0

Severity: MEDIUM

Problem Type: CWE-312,CWE-200,

Products Affected

Vendor Product Version
pifzer plum_a+_infusion_system_firmware *
pifzer plum_a+3_infusion_system_firmware *
pifzer symbiq_infusion_system_firmware *
CVE-2015-3953 HIGH

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

CVSS 2.0

Severity: HIGH

Problem Type: CWE-259,CWE-798,

Products Affected

Vendor Product Version
pifzer plum_a+_infusion_system_firmware *
pifzer plum_a+3_infusion_system_firmware *
pifzer symbiq_infusion_system_firmware *
CVE-2015-3954 HIGH

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

CVSS 2.0

Severity: HIGH

Problem Type: CWE-285,CWE-285,

Products Affected

Vendor Product Version
pifzer plum_a+_infusion_system_firmware *
pifzer plum_a+3_infusion_system_firmware *
pifzer symbiq_infusion_system_firmware *
CVE-2015-3956 HIGH

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

CVSS 2.0

Severity: HIGH

Problem Type: CWE-345,CWE-345,

Products Affected

Vendor Product Version
pifzer plum_a+_infusion_system_firmware *
pifzer plum_a+3_infusion_system_firmware *
pifzer symbiq_infusion_system_firmware *