MidnightBSD

Advisories for swisslog-healthcare

CVE-2021-37160 HIGH

A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. There is no firmware validation (e.g., cryptographic signature validation) during a File Upload for a firmware update.

CVSS 3.x

Source Score Severity Vector Exploitability Impact
nvd@nist.gov 9.8 CRITICAL CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H 3.9 5.9

CVSS 2.0

Severity: HIGH

Problem Type: CWE-347,

Products Affected

Vendor Product Version
swisslog-healthcare hmi-3_control_panel_firmware *
CVE-2021-37161 HIGH

A buffer overflow issue was discovered in the HMI3 Control Panel contained within the Swisslog Healthcare Nexus Panel, operated by released versions of software before Nexus Software 7.2.5.7. A buffer overflow allows an attacker to overwrite an internal queue data structure and can lead to remote code execution.

CVSS 3.x

Source Score Severity Vector Exploitability Impact
nvd@nist.gov 9.8 CRITICAL CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H 3.9 5.9

CVSS 2.0

Severity: HIGH

Problem Type: CWE-120,

Products Affected

Vendor Product Version
swisslog-healthcare hmi-3_control_panel_firmware *
CVE-2021-37162 HIGH

A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. If an attacker sends a malformed UDP message, a buffer underflow occurs, leading to an out-of-bounds copy and possible remote code execution.

CVSS 3.x

Source Score Severity Vector Exploitability Impact
nvd@nist.gov 9.8 CRITICAL CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H 3.9 5.9

CVSS 2.0

Severity: HIGH

Problem Type: CWE-120,

Products Affected

Vendor Product Version
swisslog-healthcare hmi-3_control_panel_firmware *
CVE-2021-37163 HIGH

An insecure permissions issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus operated by released versions of software before Nexus Software 7.2.5.7. The device has two user accounts with passwords that are hardcoded.

CVSS 3.x

Source Score Severity Vector Exploitability Impact
nvd@nist.gov 9.8 CRITICAL CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H 3.9 5.9

CVSS 2.0

Severity: HIGH

Problem Type: CWE-798,

Products Affected

Vendor Product Version
swisslog-healthcare hmi-3_control_panel_firmware *
CVE-2021-37164 HIGH

A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. In the tcpTxThread function, the received data is copied to a stack buffer. An off-by-3 condition can occur, resulting in a stack-based buffer overflow.

CVSS 3.x

Source Score Severity Vector Exploitability Impact
nvd@nist.gov 9.8 CRITICAL CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H 3.9 5.9

CVSS 2.0

Severity: HIGH

Problem Type: CWE-787,

Products Affected

Vendor Product Version
swisslog-healthcare hmi-3_control_panel_firmware *
CVE-2021-37165 HIGH

A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. When a message is sent to the HMI TCP socket, it is forwarded to the hmiProcessMsg function through the pendingQ, and may lead to remote code execution.

CVSS 3.x

Source Score Severity Vector Exploitability Impact
nvd@nist.gov 9.8 CRITICAL CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H 3.9 5.9

CVSS 2.0

Severity: HIGH

Problem Type: CWE-120,

Products Affected

Vendor Product Version
swisslog-healthcare hmi-3_control_panel_firmware *
CVE-2021-37166 HIGH

A buffer overflow issue leading to denial of service was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. When HMI3 starts up, it binds a local service to a TCP port on all interfaces of the device, and takes extensive time for the GUI to connect to the TCP socket, allowing the connection to be hijacked by an external attacker.

CVSS 3.x

Source Score Severity Vector Exploitability Impact
nvd@nist.gov 7.5 HIGH CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H 3.9 3.6

CVSS 2.0

Severity: HIGH

Problem Type: CWE-120,

Products Affected

Vendor Product Version
swisslog-healthcare hmi-3_control_panel_firmware *
CVE-2021-37167 HIGH

An insecure permissions issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. A user logged in using the default credentials can gain root access to the device, which provides permissions for all of the functionality of the device.

CVSS 3.x

Source Score Severity Vector Exploitability Impact
nvd@nist.gov 9.8 CRITICAL CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H 3.9 5.9

CVSS 2.0

Severity: HIGH

Problem Type: CWE-269,

Products Affected

Vendor Product Version
swisslog-healthcare hmi-3_control_panel_firmware *